Description
The XC-572 Source Sampling Console is the metric version of the XC-523. It is a full-featured, compact, and lightweight Isokinetic Source Sampler. It has all the great features you want and the reliability you need. Field set up is easy with non-reversible external pump connections and the industry standard 4-pin electrical connector. (External sample pump sold separately).
Console allows operators to monitor gas velocities, temperatures, pressures, and sample flow rates for maintaining isokinetic sampling conditions.
All applicable components are rated 120V / 60Hz or 240V/50Hz.
Features:
- Precision gas meter with orifice
- Rugged and lightweight roto-molded linear low-density polyethylene (LLDPE/LMDE) case with two molded-in ergonomic carrying handles
- Low-cost manual console
- Easy operation and maintenance
- Rugged, lightweight, stackable case
- Removable 19-inch front panel
- Carry strap for easy transport
- Non-reversible sample pump connections
- Color-coded thermocouple jacks
- Bright, easy-to-read temperature displays
- Easy-leveling dual-column manometer
- Stainless steel quick connects and fittings
- Convenient layout for manual data recording
Specifications:
Dry Gas Meter:
Model SK25EX
Multi-chamber positive displacement
Qmax: for air 4.1m3/h at 220 Pa
Qmin: 0.26 L/min resolution 0.002 Liter
Cyclic volume: 0.7L
41 LPM max
Volume display via mechanical index
Accuracy: 1.5% at volumes greater than 10 liters
Type-K thermocouple for exit temperature
Manometer:
Inclined/Vertical manometer for determining stack velocity and sample flow rate
Inclined range of 0-26 mm H2O with a resolution of 0.2 mm
Vertical range of 26-250 mm H2O with 2 mm divisions
Fabricated from solid acrylic, precision-bored, accurate to +/- 1%
Convenient self-sealing chrome plated brass quick connects with Viton O-rings
Supplied with plastic plugs for a positive secondary seal
Temperature Display:
Panel mount digital thermometer
Bright red LED display
Type-K thermocouple with Range -105°C to 1372°C
Separate 7 channel selector switch
Temperature Controllers- Probe and Oven:
Programmable thermocouple type
Range 0°C to 1200°C
Digital setting by 3 keys with lock function
NEMA 4X/IP66 equivalent waterproof faceplate
Large red 4-digit LED display
Alarm output capable
Vacuum Gauge: dual-scale, 0-30inHg / 0-100kPa
Sample Pump Connections: 3/8-inch non-reversible, stainless-steel quick connects
Umbilical Connections:
Sample inlet: 1/2-inch stainless steel quick connect
Pitot lines: 1/4-inch stainless steel quick connect
Electrical: 5 Pin female bayonet PT series Amphenol circular connector
Thermocouples: quantity 6 type-K inputs, standard size
Power: 120VAC at 60Hz standard (240VAC at 50Hz optional), IEC C14 male receptacle
Dimensions: 23” x 21” x 12” (58cm x 53cm x 30cm), 19-inch panel
Weight: 40 lbs (18kg)